MEANING OF ESTIMATION VULNERABILITY ON COMPUTERIZED MULTI-METERS

Section: Articles Published Date: 2021-11-11 Pages: 68-71 Conference Proceedings Volume: Results of Modern Scientific Research and Development

Authors

  • Dr.Vikram Seth Calibration Lab, Amara Raja Batteries Ltd. Renigunta- Kadapa Road, Karakambadi-517520, Andhra Pradesh, India.
PDF

Abstract

This paper presents the alignment results with vulnerabilities of three computerized multi-meters to be specific A, B and C that are utilized in different confined conditions to comprehend the adjustment execution of the meters in both in ac and dc possibilities. The vulnerability (Ue) values are determined for both ac and dc possibilities at characterized resistance at set focuses 3.8, 38 and 380V individually. The adjustment execution of the meters A, B and C are viewed as in the request for A>B>C. The meaning of vulnerability esteems are associated by bookkeeping the adjustment history, use design and the restricted conditions.

Keywords

Adjustment, vulnerability esteem, resistance, advanced multi-meters, confined conditions